Surgical instrument care and sterilisation

How to clean, inspect, sterilise and store reusable stainless steel instruments so they stay sharp, smooth and free of corrosion for years.

  • Applies toReusable stainless steel instruments
  • WaterDemineralised for final rinse
  • LubricantWater-soluble, steam-permeable
  • SterilisationValidated steam cycle

Always follow the instructions for use (IFU) supplied with the instrument and your facility’s validated reprocessing procedures. This guide gives general good practice for reusable stainless steel instruments. It does not replace the IFU, local regulations or your steriliser manufacturer’s instructions.

Six steps for every cycle

Most instrument damage happens in the first hour after surgery, and in the final rinse and drying. Get those right and instruments last much longer.

  1. Treat at point of use

    Wipe off visible soil straight after use and keep instruments moist, for example with a damp towel or pre-cleaning spray. Do not let blood or tissue dry on them, and never soak in saline.

  2. Clean

    Open hinged instruments and take apart those designed to come apart. Clean in a washer-disinfector, ultrasonic bath or by hand, with a detergent recommended for surgical instruments and soft nylon brushes. Clean cannulas and lumens with a brush of the right size.

  3. Rinse and dry

    Rinse thoroughly to remove all detergent, using demineralised or deionised water for the final rinse to avoid spots. Dry completely, including joints and lumens.

  4. Inspect and lubricate

    Check under good light for residue, corrosion, cracks, burrs and blunt edges. Test that jaws meet, ratchets hold and scissors cut cleanly. Lubricate hinges and box joints with a water-soluble instrument lubricant.

  5. Pack and sterilise

    Pack instruments open or on the first ratchet click, without overloading trays. Sterilise with a validated steam cycle as set out in the IFU, for example 134 °C for at least 3 minutes in a pre-vacuum steriliser, and let packs dry fully before handling.

  6. Store

    Store sterile packs in a clean, dry area away from chemicals and direct sunlight. Keep sharp tips and fine jaws protected so they are not bent or blunted.

What damages instruments

Stainless steel resists corrosion, but it is not indestructible. Avoid these and you avoid most rust, pitting and stiff joints.

  • Letting blood, tissue or saline dry on the surface
  • Bleach, chlorine or other chloride solutions
  • Steel wool, wire brushes or abrasive powders
  • Hard tap water for the final rinse
  • Mixing stainless steel with other metals in the ultrasonic bath
  • Oil or silicone lubricants that block steam
  • Overloaded trays and heavy instruments on top of delicate ones
  • Wet packs after sterilisation

Instruments that need extra care

Scissors

Clean with blades open. Keep cutting edges from touching other instruments, and send for sharpening before they tear tissue.

Micro and ophthalmic

Handle one at a time, protect tips with caps or silicone mats, and clean in their own tray.

Ratchet instruments

Sterilise open or on the first click. Storing them closed tight strains the joint and can cause cracks.

Single-use instruments

Designed for one procedure only. Do not clean, resterilise or reuse them; dispose of them safely.

Care questions

Why do stainless steel instruments get rust spots?

Usually from residues, not the steel itself: dried blood, saline, chlorine or bleach, hard-water minerals, or rust transferred from damaged instruments in the same load. Clean promptly, rinse with demineralised water, dry fully and remove any corroded instruments from circulation.

Can I use oil to lubricate hinges?

No. Use a water-soluble, steam-permeable instrument lubricant (often called instrument milk). Oils and silicone sprays block steam from reaching the surface during sterilisation.

What causes stains or discolouration after the autoclave?

Common causes are hard or contaminated water in the steriliser, detergent residues that were not rinsed off, and wet packs. Check water quality and rinsing first.

Can single-use instruments be reprocessed?

No. Single-use instruments are designed for one procedure only. Dispose of them safely after use.

Where do I find the instructions for a specific instrument?

Follow the instructions for use supplied with the instrument. If you need a copy for any of our instruments, ask us and we will send it.

Need instructions for use or replacements?

Ask us for the IFU of any of our instruments, or send a list of instruments you need to replace.

Need an instrument that isn’t in our catalogue?

Send us a drawing, a photo or a sample. We manufacture to your specification, sterile or non-sterile, in your own packaging, and ship worldwide with full export documents.

  1. 1

    Send your drawing or sample

    Share a technical drawing, a photo or a physical sample, with quantity and packaging needs.

  2. 2

    Approve quotation and prototype

    We quote price and lead time, then make a sample for your approval.

  3. 3

    Production and worldwide shipping

    Forged, finished and inspected in our Sialkot factory, shipped with CE documentation.