Short answer: Reusable surgical instruments go through the same cycle after every use: pre-clean at the point of use, transport, clean, rinse, dry, inspect, lubricate, pack, sterilise and store. Every step matters. If cleaning fails, sterilisation fails too.
This guide explains each step in plain language, with tips from experienced reprocessing teams, common mistakes to avoid and links to official guidance.
This guide is general information. Always follow your local infection control policy, your national guidelines and the manufacturer’s instructions for use (IFU).
The instrument reprocessing cycle

- Pre-clean at point of use
- Transport in a closed container
- Clean (manual, ultrasonic or washer-disinfector)
- Rinse with good-quality water
- Dry completely
- Inspect and test function
- Lubricate joints
- Pack (wrap, pouch or container)
- Sterilise
- Store dry and protected
Why the order matters
Sterilisation kills microorganisms on surfaces it can reach. Dried blood, tissue or detergent residue shields microorganisms from steam. That is why international guidance treats cleaning as the most important step.
Step 1: Pre-clean at the point of use
Straight after the procedure:
- Wipe visible blood and tissue off instruments with a damp swab
- Open hinged instruments
- Keep instruments moist (for example with a damp towel or a foam/gel product made for this purpose) until cleaning
- Separate sharps safely
Tip: Dried blood is the number one cause of stains and failed cleaning. The faster instruments reach cleaning, the better.
Step 2: Transport
Move used instruments in a closed, labelled container to the cleaning area. This protects staff and prevents drying out.
Step 3: Cleaning
There are three main methods. Many departments use more than one.
| Method | Best for | Notes |
|---|---|---|
| Manual cleaning | Delicate or complex instruments, pre-cleaning | Use soft brushes, clean under the water surface to avoid splashing |
| Ultrasonic cleaning | Joints, serrations, hard-to-reach areas | Do not put certain delicate or optical items in unless the IFU allows |
| Washer-disinfector | Routine, high-volume cleaning | Validated, repeatable, includes thermal disinfection |
Detergent: Use a detergent designed for surgical instruments, at the correct dose and temperature. Enzymatic and mildly alkaline detergents are common. Avoid household products.
Joints and lumens: Clean box locks, ratchets, serrations and channels carefully. These are where soil hides.
Step 4: Rinsing
Rinse off all detergent. The final rinse should ideally use demineralised or reverse-osmosis water, which prevents white spots and stains.
Step 5: Drying
Dry thoroughly, including joints and lumens. Water left behind can cause spots, corrosion and wet packs after sterilisation.
Step 6: Inspection and function testing
Check every instrument, ideally under good light and magnification.
- Visibly clean: no blood, tissue or residue
- No cracks, especially at box locks
- Jaws and tips align
- Ratchets hold and release correctly
- Scissors cut cleanly (test on a test material)
- No rust, pitting or damaged coatings
Note: Remove damaged instruments from the set and send them for repair or replacement. One rusty instrument can stain others in the same load.
Step 7: Lubrication
Apply a water-soluble instrument lubricant (“instrument milk”) to joints and moving parts. Never use oils or silicone sprays: they block steam and trap dirt (see What is instrument milk, and do I need it?).
Step 8: Packing
| Packaging | Use |
|---|---|
| Sterilisation wraps | Trays and sets |
| Paper/film pouches | Single instruments, small sets |
| Rigid containers | Sets used often, protects instruments |
- Pack ratchets open or on the first tooth only
- Protect delicate tips with caps
- Add a set list and indicators as your policy requires
- Label with contents, date and load number
Step 9: Sterilisation
Steam sterilisation (autoclaving) is the most common method for stainless steel instruments. Widely used cycles include about 121°C for a longer exposure time or about 134°C for a shorter exposure time. The correct cycle depends on your steriliser, your packaging and the instrument IFU.
Monitoring usually includes:
- Physical monitoring: Temperature, pressure and time printouts
- Chemical indicators: Show the pack was exposed to the process
- Biological indicators: Test that spores were killed, at intervals set by your policy
Step 10: Storage
- Store packs in a clean, dry, dust-free area
- Do not stack heavy trays on pouches
- Use stock in date order
- Check packs before use: discard torn, wet or damaged ones
Common mistakes and how to avoid them

- Dried blood: Pre-clean straight after use and keep instruments moist
- Saline soaking: Saline causes pitting; use water or a pre-clean product
- Bleach: Chlorine damages stainless steel; never use it on instruments
- Overloading: Overcrowded loads clean and sterilise poorly
- Closed ratchets: Sterilise open or on the first click to avoid cracks
- Wet packs: Dry instruments fully and let loads cool before handling
Worked example: a minor surgery set in a small clinic
A small clinic uses one minor surgery set (12 instruments) several times a day.
- After each patient the nurse wipes the instruments and places them open in a closed box with a pre-cleaning spray.
- At lunch and at the end of the day, instruments are cleaned in an ultrasonic cleaner, rinsed with demineralised water and dried.
- Each item is checked under a lamp with a magnifier. A needle holder with a loose ratchet is set aside.
- Joints are lubricated, the set is packed in a pouch with a set list and indicator, and sterilised in the clinic’s autoclave.
- Packs are dated and stored in a closed cabinet.
The clinic keeps three sets in rotation so one is always ready.
Daily reprocessing checklist

- Instruments pre-cleaned at point of use
- Correct detergent dose and temperature
- Final rinse with demineralised water
- Every item inspected and tested
- Joints lubricated
- Ratchets open when packed
- Cycle printout and indicators checked
Proof box
Proof: ISO 17664 requires manufacturers to provide validated reprocessing information, and ISO 15883 sets requirements for washer-disinfectors. Steam sterilisation processes are covered by ISO 17665.
Key takeaways
- Cleaning is the most important step: you cannot sterilise dirt
- Use the right detergent and good water
- Inspect every instrument, every time
- Lubricate, pack correctly and sterilise with a validated cycle
- Remove damaged instruments early
Resources
- CDC: Guideline for Disinfection and Sterilization in Healthcare Facilities
- AKI (Instrument Reprocessing Working Group): “Proper maintenance of instruments” brochures
- WHO: Medical devices
- Related articles: rust and stains (see Why do surgical instruments rust or stain?), instrument milk (see What is instrument milk, and do I need it?), sterilisation methods for single-use products (see Sterilisation methods for single-use instruments: EO, gamma and steam)